Inflatable life preservers are
subject to general deterioration due to aging. Experience has indicated that
such equipment may be in need of replacement at the end of 5 years due to
porosity of the rubber-coated material. Wear of such equipment is accelerated
when stowed on board aircraft because of vibration which causes chafing of
the rubberized fabric. This ultimately results in localized leakage. Leakage is
also likely to occur where the fabric is folded because sharp corners are
formed. When these corners are in contact with the carrying cases, or with
adjacent parts of the rubberized fabric, they tend to wear through due to
vibration.
Life preservers should be
inspected in accordance with the manufacturer's specification, unless climate,
storage, or operational conditions indicate the need for more frequent inspections.
Life preservers should be inspected at 12-month intervals for cuts,
tears, or other damage to the rubberized material:. Check the mouth valves and
tubing for leakage, corrosion, and deterioration: Remove. the CO2
cylinder and check the discharge mechanism by operating the lever to ascertain
that the pin operates freely. Check the gaskets and valve cores of the
cylinder container and the pull cord for deterioration. If no defects are
found, inflate the preserver with air to a 2 psi pressure and allow to stand
for 12 hours. If the preserver still has adequate rigidity at the end of that
time, deflate and fit with C02
cylinders having weights not less than that indicated on them by the
manufacturer. All cylinders made in accordance with joint Army-Navy Specification MIL -C-00601
D are' so stamped and have a minimum permissible weight stamped on them. The
use of such CO2 cylinders is recommended. Having fitted the
preserver with an adequately-chaiged cylinder, mark the preserver to indicate
the date of inspection and patch it to the container. It is recommended that
the aforementioned procedure be repeated every 12-month period, utilizing the
C02 cartridge. for inflation. Carbon dioxide permeates the
rubberized fabric at a faster rate than air and will indicate if the porosity
of the material is excessive. The following checks and inspections should be
completed:
a. Check for abrasions,
chafing, and soiling across folded cell areas and around metal parts. Condemn
the life preserver when unsuitable conditions are found.
b. Check for separation of cell fabric and loose attachments along the
edges of patches and sealing tapes. Repair if practicable.
c. Check for deterioration in areas where oil and grease are noted. Condemn
deteriorated cells. If deterioration is not noted, clean the areas with mild
soap and water and rinse with clear water.'
d. Inspect the snaps and/or buckles to ensure proper operation.
e. Inspect the instruction panel for readability.
f. Inspect all stitching for gaps, pulls, and tears.
g. Visually inspect the cell containers for snags, cuts, loose stitching, and oil and
grease spots. Repair or replace as necessary.
h. Inspect the hardware for rusted or broken parts and cotter pins for
damage. Ensure that pins are smooth and free of burrs.
i. Check the inflator discharge lever for proper operation. Move the inflator discharge lever slowly through a
normal cycle of operation to ensure freedom of 'operation and to make certain
that the piercing pin has sufficient movement to discharge the COZ cylinder. The point of the pin should move past the surface of the gasket in the
inflator. In the un-operated position, the end paint should be slightly below
the gasket surface.
j. Check the installation of the inflator stem gaskets and check the stem
caps for tightness.. Ensure that the inflator is centred on the stem.
k. Check rescue light. Inspect and test.
(1) Replace the
battery if it shows any signs of encrustation.
(2) Inspect for
proper installation and physical condition of the lamp, wire, and battery.
(3) Check the
light assembly for proper operation and water insulation and flotation.
(4) Pull the
sealing plug (where applicable) from the battery. Let water flow through the
open ports. Make sure the battery is activated and power is supplied to the
light.
(5) Fill out
the inspection record and serviceable parts tag. Attach to the vest.
l. Deflate the life preserver and repack in
container and secure.
m. The accessories listed below will be required for all life preservers:
(1) One
Recognition Light: Remove when returning to serviceable or reparable storage.
Remove for replacement of defective light, repair, or salvage of preserver.
(2) One
Recognition Light Battery : Remove when
returning to serviceable or reparable storage.
n. Record the inspection data on data cards.
o. Life preserver inspected and found sea
worthy. Include the inspector's
signature.
p. Inspection record. Upon completion of 12-month inspection and tests,
each flotation cell will be marked to indicate the date the inspection was
accomplished. The inspection stencil will consist of 1/8-inch letters and numerals
and will be applied to the patches on the cells (example: 4/3/97 ). To facilitate determination of the next
12-month inspection period, enter the date it is due in the blank beside the
word inspect on the inspection data card provided in the inspection data pocket
on the cell container. Repack, close, and seal the container.
REPAIR OF LIFE PRESERVERS
Leaks may be disclosed by immersion in soapy water. Repair leaks by the use of patches in accordance with the recommendations of the manufacturer. Clean corroded metal parts and replace missing or weakened lanyards. Life preservers which do not retain sufficient rigidity after the 12-hour period, because of general deterioration and: porosity of the fabric, are beyond economical repair and should be replaced.
No comments:
Post a Comment