Oxygen
System Servicing, Inspection And Maintenance Practices
INTRODUCTION
Oxygen system requires some careful
maintenance activities. This week highlight different type of servicing, inspection and maintenance procedures.
Oxygen
System Servicing: GENERAL
Care
and attention to detail is the mark of professional aviation maintenance, and
nowhere is this characteristic more important than when servicing aircraft
oxygen systems. Compressed gaseous oxygen demands special attention because of
both its high storage pressure and its extremely active chemical nature.
When
possible, all oxygen servicing should be done outdoors, or at least in a well
ventilated area of the hangar. Oxygen systems having removable or portable
supply cylinders should have these containers removed from the operation, and
all electrical work within the aircraft should be suspended during the
servicing. In all cases the manufacturer's service information must always be
used while performing service, maintenance or inspection on aircraft oxygen systems.
Servicing Gaseous Oxygen Systems
Leak Testing Gaseous Oxygen
Systems: Leaks should be searched out by using a
special leak detector material which is a form of non-oily soap solution.
Spread this solution over every fitting and
at every place a leak could possibly occur, and the presence of bubbles will
indicate a leak. If a leak is found, release the pressure from the system, and
check the fittings for proper torque. It is especially true of flare less
fittings where over tightening can intensify a leak. If the fitting is properly
torqued and still leaks, remove the fitting and examine all of the sealing
surfaces for indication of damage. It may be necessary to replace the fitting
and reflare the tube or install a new flare less fitting.
Draining The Oxygen System: Draining
of the oxygen system should normally be done after the high pressure bottle has
been removed or isolated from the system. Either outdoors or in a well
ventilated hangar, open the aircraft's doors and windows, then bleed the
system's pressure off by opening the appropriate fitting to allow the oxygen
pressure to bleed off. Normally a system will require purging after the system
has been drained. All the safety precautions mentioned later in this chapter
should be followed during any oxygen draining procedure.
Filling
An Oxygen System: Fixed base operators who do a
considerable amount of oxygen servicing will usually have one of the larger
oxygen servicing carts such as the one seen in Figure 7.1. This cart carries six large cylinders, each holding
approximately 250 cubic feet of aviators breathing oxygen. A seventh cylinder,
facing the opposite direction and filled with compressed nitrogen, is normally
carried to charge hydraulic accumulators and de-icer cylinders. Fittings on the
nitrogen cylinders are different from those on the oxygen cylinders, to
minimize the possibility of using nitrogen to fill the oxygen system, or of
servicing the other systems with oxygen.
If you ever
fill a low-pressure oxygen system from a high-pressure supply, be sure the
proper regulator is installed and the output pressure is adjusted to that
required for the system.
Various
manufacturers of oxygen equipment use different types of connections between
the supply and the aircraft, and a well equipped service cart should have the
proper adapters. These adapters must be kept clean and protected from damage.
Never improvise when adapting a supply cart to the aircraft. Leakage during the
filling operation is not only costly, but it is hazardous as well.
Before filling
any aircraft oxygen system, be sure that all of the cylinders are of the
approved type, and that they have all been hydrostatically tested within the
required time interval.
No oxygen
system should be allowed to become completely empty. When there is no pressure
inside the cylinder, air can enter, and most air contains water vapour. When
the water vapour is mixed with the oxygen and expanded through the small
orifices in the system, the water is likely to freeze and shut off the flow of
oxygen to the masks. Water in a cylinder can also cause it to rust on the
inside and weaken it so it will fail with catastrophic results. A system is
considered to be empty when the pressure gets down to 50 to 100 psi. If the
system is ever allowed to get completely empty, the valve should be removed and
the cylinder cleaned and inspected by an FAA-approved repair station.
When filling an
airplane from a large supply cart, start with the cylinder having the lowest
pressure. The pressure should be written on the container with chalk or a
record kept with the cart. Momentarily crack the valve on the cylinder and
allow some oxygen to purge all of the moisture, dirt and air from the line;
then connect the line to the aircraft filler valve and slowly open the valve on
the cylinder. Most all filler valves have restrictors that prevent too high a
flow rate into the cylinder. When the pressure in the aircraft system and that
in the cylinder with the lowest pressure stabilizes and there is no more flow,
mark this pressure on the cylinder with chalk and close the
cylinder valve. Slowly open the valve on the cylinder having the next lowest
pressure and allow oxygen to flow into the system until it again stabilizes. Continue this procedure until the pressure in the aircraft
system is that which is required.
The
ambient temperature determines the pressure that should be put into the oxygen
system, and a chart similar to the one in should be used to determine the pressure needed. For example, if the
ambient temperature is 90 °F and you want a stabilized pressure in the system
of 1,800 psi, you should allow the oxygen to flow until a pressure of 2,000 psi
is indicated on the system pressure gauge. When the oxygen in the system drops
to its standard temperature of 70 °F, the pressure should stabilize at 1,800
psi. If the ambient temperature is low, you must stop filling the system at a
lower pressure, because the oxygen will expand and the pressure will rise when
it warms up to its standard temperature.\
Purging A Gaseous Oxygen
System: If
the oxygen system has been opened for servicing, you should purge it of any
air that may be in the lines. To purge a continuous flow system, plug masks
into each of the outlets, turn on the oxygen supply valve, and allow the oxygen
to flow through the system for about ten minutes. Diluter demand and pressure
demand systems may be purged by placing the regulators in the EMERGENCY position
and allowing the oxygen to flow through them for about ten minutes. After the
system has been thoroughly purged, fill the cylinders to the required
pressure.
No comments:
Post a Comment